Specifications
TYMPANOMETRY MODE
Probe frequency, -intensity: 226 Hz ± 1%, 69 dB HL
Optional: 1000 Hz ± 1%, 69 dB HL
Pressure range: -400 to +200 daPa
Volume range: 0.1 to 8.0 ml
Accuracy: ± 5% or ± 10 daPa
Test time: < 3 seconds
REFLEX MODE
Test frequencies: 500, 1000, 2000, 4000 Hz ± 3%
Test Method: Ipsilateral and contralateral
Intensities Reflexes: 70 ... 100 dB HL
Intensity setting: Automatic or single intensity
OPT. HIGH FREQUENCY
- Tymp 1 kHz;
- Tymp 1 kHz + auto reflex;
- Tymp 1 kHz + reflex 80 dB BB
GENERAL
Memory: 1 curve per ear and test
Internal Memory: > 1000 measurements
Display: Full color graphic LCD-Display
Dimensions / Weight: W x D x H: 8 x 30 x 7 cm / 15oz
CRADLE
Dimensions: W x D x H: 15 x 29.5 x 11 cm
Power: 100 - 240 VAC, 50/60 Hz
REFLEX DECAY MODE
Test Time: 0 to 12 seconds
Test Method: Ipsilateral and contralateral
ETF INTACT
Pressure Range: -400 to +200 daPa
ETF PERFORATED
Probe frequency: 226 Hz ± 1%
Test Time: 0-30 sec. auto tone present